YOU CAN DO THE ACTIONS LISTED BELOW ONLY IF YOU WERE DIRECTLY INSTRUCTED BY NMR CENTER STAFF .
NOTE: If your solvent is toluene-d8 (Tol) and 'atma' fails with an error, follow these instructions: System will not tune when lock solvent is toluene-d8
Tuning system needs to know what nucleus you are going to work with. Follow these steps:
Manual tuning mode allows you to see what channels are tuned, inspect the result of tuning and adjust if necessary
NOTE: The automatic tuning result is completely acceptable for most routine work. However, for best sensitivity you may want to adjust Tuning and Matching manually to make the resonance sharper.
Probe is now tuned and ready for experiments.
Next step is to lock spectrometer on your solvent:
Wait for the lock system to finish locking (see the status messages)
Shimming is required to achieve good resolution in NMR spectra.
Using Temperature Control
IMPORTANT NOTE: If you need to use temperature OUTSIDE of 20-40 Deg C range,
you must be trained on the "Variable Temperature" protocol: contact NMR center staff to schedule training.
For all Variable Temperature tutorials and instructions see Variable-temperature NMR operation
The instrument is equipped with manual BCU-1 refrigerated chiller and liquid N2 chiller.